2.20.9

New Features

  • Added support for firmware v1.150 (104) for RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRbrevio3, RBRquartz3

  • Updated Quick Start Guide (QSG) for RBRquartz3 BPR|zero

  • Updated RBR Instrument Guide (RIG) for RBRquartz3 BPR|zero

  • Updated RBR Instrument Guide (RIG) for RBR CT and CTD instruments

  • Updated RBR Instrument Guide (RIG) for Sensors, Compact instruments, and Standard instruments

  • Updated RBR Instrument Guide (RIG) for RBRlegato3

  • Added factory support for TRI and QUAD part numbers

  • Factory configures all instruments via configuration files. If not present will then look at tokenizing part number

  • Added support for STID16, STRID17, and STRID18 sensor packs

Improvements

  • RSK-10251 - Updated default pressure compensation coefficient for ODO concentration

  • RSK-10542 - Improved optical PAR calibrations with hardcoded C0 coefficients

  • RSK-10684 - We now handle normalizing the quadrante data in the downloaded RSK file

  • RSK-10725 - Updated Gordon and Garcia solubility constants

  • RSK-10746 - Temperature references are ordered coldest to warmest on certificates

  • RSK-10767 - Added lisocl2 battery option for RBRduo3 T.T

  • RSK-10801 - Updated Ccal resistance reference values

  • RSK-10837 - Added 750dbar option to legacy pressure calibration routine

  • RSK-10842 - Updated settling/read times for RBRcoda T.ODO angled fast foil

  • RSK-10868 - Added channel calibration range parameter to differentiate it from maximum range

  • RSK-10894 - Allow Transmissometer channel to be user calibrated

  • RSK-10904 - Added support for more RBRtridente channel types

  • RSK-10907 - Updated Pcal conductivity stability threshold

Bug fixes

  • RSK-10617 - Fixed NPE on generic RBRtridente gain calibration process

  • RSK-10625 - Fixed calibration equation sizing on Chlorophyl and Fluorometer certificates

  • RSK-10635 - Autocommit was missing on some SQL calls preventing recalculation of samples following coefficient updates

  • RSK-10651 - Fixed lock monitor so we can update firmware on FE cards

  • RSK-10672 - Firmware upgrade dialog shows a locked null version on virgin tridente

  • RSK-10737 - Fixed an offset error when downloading binary chunks using variable chunk sizes

  • RSK-10753 - Fixed NPE on Tcal batch calibration process

  • RSK-10891 - Configuration table for Wetlabs ECO puck sensor on FE-serial3 had some incorrect settings which prevented communication

  • RSK-10928 - Some sensors can display 'infinity' for sleep power under some sampling conditions