O2 - compensated for pressure and salinity
Let us consider the RBRconcerto³ C.T.D.ODO, where the ODO channel is an integrated RBRcoda3 T.ODO. The RBRcoda3 T.ODO transfers both the foil temperature and the dissolved oxygen concentration (not compensated for salinity). The RBRconcerto³ calculates the concentration compensated for salinity and pressure first:

where
-
O2corr is the corrected O2 concentration, compensated for salinity and pressure,
-
O2unc is the uncompensated O2 concentration returned by the RBRcoda3 T.ODO,
-
C0 and C1 are corrections and scaling factors for the uncompensated O2 concentration,
-
C2 is a correction factor for pressure,
-
S is the salinity in PSU,
-
P is atmospheric pressure in dbar,
and

where T is the water temperature (in °C), and

which correspond to Garcia and Gordon coefficients.