Generation² twist activation troubleshooting
For Generation2 instruments (pre 2018).
If the twist activation is not functioning on a Generation2 instrument, verify that the twist activation feature is installed and enabled on it. Then, follow the steps in Twist activation troubleshooting.
An additional step that can be taken for these instruments, is probing the twist activation magnet present in the cover above the battery carriage. The magnet fills one of the holes in this cover. When twist activation is on, this magnet can be probed with a kitchen magnet to check that the twist activation is working. If there is no vibration when the setting is enabled, this confirms that there is an issue with the twist activation.
Ensure that the end-cap matches the instrument and that the magnets are securely in place, both on the end-cap and in the battery cover of the instrument.

If these are all in place and the magnets still do not align, you may need to return the instrument to RBR for repair. Please contact our support team to obtain a return merchandise authorisation code (RMA) and review the detailed shipping information on the RBR website.