Notice
This notice pertains to the following instruments:
RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRquartz3, RBRbrevio3, RBRsolo3, and RBRduet3
The notice pertains to the following sampling regimes:
Averaging, Bursting, Tide, or Wave
Recently it was discovered that Generation3 instruments using an intermittent sampling mode (bursting, averaging, tides, and waves) would present incorrect deployment autonomy estimates - the estimates are shorter than actual behaviour. The cause of the discrepancy is due to an incorrect sleep value used in the autonomy engine. Deployments where sampling occurs only for a small fraction of the burst interval are affected most.
Please contact [email protected] for any questions or concerns regarding this notice.
Improvements
-
RSK-6268 - Decimal places reported on calibration certificates have been reduced
-
RSK-6288 - Mouse cursor changes when in annotation mode
-
RSK-6635 - Simulated deployments life doesn’t match physical instrument due to memory size estimation
-
RSK-6642 - Fixed potential deadlock in Development kit configuration
-
RSK-6646 - Display channel error code in data table
-
RSK-6648 - Added calibration stage to the channel calibration object
-
RSK-6665 - Pulled generation 3 part code descriptions into configuration files
-
RSK-6694 - Restrict instrument derived channels when fast32 feature is included on instrument
-
RSK-6700 - Decreased time required to reprocess tide and wave data
-
RSK-6701 - Reprocessing data no longer blocks the UI
-
RSK-6708 - Fixed increased CPU usage caused by UDP server
-
RSK-6710 - Fixed potential deadlock issue on the UI when closing Ruskin
-
RSK-6713 - Added default instrument settings to simulated instruments to display on the parameters tab
-
RSK-6726 - Fixed slow transition through time breaks in plot
-
RSK-6730 - Opening RSK file now sets focus in the Navigator
Bug Fixes
-
RSK-6538 - Simulated RBRduet T.ODO displays incorrect sampling rates
-
RSK-6719 - Ruskin is trying to read streaming data from a Wi-Fi port
-
RSK-6720 - Disposing of the instrument editors throws exceptions
-
RSK-6722 - Ruskin left a hanging application in the background
-
RSK-6723 - RBRquartz Q live plot is not being created
-
RSK-6731 - Reprocessing tide data doesn’t work